Online and memory probe with internal calculation

The CTD48Mc is a high quality, high accuracy memory probe for oceanographic and limnological measurement of conductivity, temperature, depth and with one optional parameter for depth up to 2000 m. The CTD48Mc is characterised in particular by its ability to directly output physical and calculated data such as salinity, sound velocity and density.

Sea & Sun Technology
Trappenkamp, DE

Description

The CTD48Mc probe is a very small and handy microprocessor controlled multiparameter memory probe for precise measurements as well as for self-contained operation in deep and shallow water with up to 4 sensors and a maximum depth down to 2000 m. It measures conductivity, temperature, pressure and one optional parameter.
The difference between the CTD48Mc and the previous version of CTD48 Probe are:

  • low weight
  • powerful micro controller
  • calibration coefficients stored inside the probe
  • output of calculated / physical data
  • powerful interfaces
  • rechargeable batteries (NO opening of the probe!)

The CTD48Mc can work in different modes. One mode is standard SST mode, as with all previous SST probes. In this mode, the data can be read, graphed and saved via an SST programme.
In another mode, physical and calculated data, such as salinity, density or speed of sound, can be readout directly as readable characters (ASCII).
The probe can be operated in the following environments:

Online operation
During measurements, the probe must be connected by a cable to a device for displaying and recording measured values. The PC with the “SDA” software application represents the device for displaying and recording measured values in this case. The probe transmits the measured values to this connected data acquisition unit, i.e. the probe does not store any measurement data in its internal data memory. Power is supplied by the internal battery or from an external source.

The online operation is implemented by the following operation mode:

  • RS-232 (factory default)
  • USB UART (factory default – for configuration and memory readout)
  • Ethernet (10/100 Mbit/s)

Memory operation
The probe can perform measurements automatically and save measurement results internally. Using a configuration cable, a software application transmits to the probe from a PC the settings that control the probe in memory operation. During the actual measurement, no connection is required between the probe and the control unit. In memory operation, the electrical supply is gener¬ally provided by the internal batteries; alternatively, power supply via a connection cable is possible. After measurement, the internally stored measured values are transmitted to a device for storage and display. In this case, the PC running the “SDA” software application is the device con¬figuring the probe and displaying and storing the measured values. The probe can also transfer the currently measured values to a device for display or storage during memory operation and using a connecting cable.

The memory operation is divided into the following operation modes:

  • Continuous mode: each data set is stored.
  • Time mode: data sets are only stored at programmable intervals with several selectable schemes.
  • Increment mode: data sets are stored at programmable depth stamps.

The CTD48Mc is equipped with an 8 channel data acquisition system with 20 bit resolution. High long-term stability and automatic self-calibration of analogue digital converter guarantee stable and precise CTD measurements for many years.

The probe power supply of the is activated by touching a reed contact with a magnetic rod. LED displays power supply status and optical control of memory access.

The supplied Standard Data Acquisition Software package “SST-SDA” includes the handling of the logging process and the display of online or recorded data with a shared graphic user interface.

Standard models:
CTD48Mc:
Online and memory probe for measurement of conductivity, pressure, temperature + 1 optional sensor

Standard sensors:

  • Conductivity (Š”)
  • Temperature (T)
  • Pressure (D)

Additional sensors:

  • Oxygen
  • Turbidity
  • pH
  • Redox (OPR)
  • Flourometer

Calculated parameters:

  • salinity
  • sound velocity
  • Kappa 25 mS/cm (electric conductivity)
  • sigma – t (potential density of a fluid)

Specifications

General

Weight in air [kg]

1.2

Diameter [mm]

48

Length [mm]

575

Housing Material

Titanium

Max. cable length [m]

2000

Observations

Min. depth increment [m]

0.01

Min. time increment [s]

0.1

Resolution of salinity [psu]

0.001

Uncertainty of salinity [psu]

0.002

Min. sound velocity [m/s]

1400

Max. sound velocity [m/s]

1600

Resolution of sound velocity [m/s]

0.05

Uncertainty of sound velocity [m/s]

0.1

Operation Characteristics

Data display on dedicated hardware

N

Method of first recording

internal /external

Display of analogue data

Y

Pressure, temperature and conductivity sensor

Pressure sensor

Y

Type of pressure sensor

piezo resistive

Resolution of pressure [% FS]

0.002

Accuracy of pressure sensor

up to 0.05

Temperature sensor

Y

Type of temperature sensor

Pt 100 4-pole

Min. temperature [Ā°C]

-2

Max. temperature [Ā°C]

36

Resolution of temperature [Ā°C]

0.0005

Accuracy of temperature measurement

0.002

Conductivity sensor

Y

Type of conductivity sensor

7-pole-cell

Max. conductivity [S/m]

70

Resolution of conductivity [10^-3 S/m]

0.0005

Accuracy of conductivity measurement

Ā± 0.002

Uncertainty of pressure [% FS]

130

Software specifications

Software name

SST-SDA

Graphical presentation

Y

Raw data editing

Y

Re-computation with raw data

Y

CTD Probe controller

Y

Recording with third party software

Y

Environmental

Max. depth [m]

2000

Max. pressure [bar]

200

Operating temperature: min. [Ā°C]

-10

Operating temperature: max. [Ā°C]

60

Data and connectivity

External recorder required

N

Direct data reading

Y

Internal memory

Y

Connectors

RS232

Internal Memory [MB]

16000

Type of internal memory

non-volatile memory

Max. number of samples for self-recording [10^3]

500000000

Dedicated sub-surface data logger

Y

Dedicated surface data logger

Y

Telemetry link to remote location

Y

Power

Min. operating power [V]

7

Max. operating power [V]

15

Battery

Y

Battery type

Alkaline / Lithium

Additonal sensors

Standard sensor

pressure, temperature, conductivity

No. of additional sensors

1

Description of additional sensors

optical oxygen (SST-DO), dissolved oxygen, pH, ORP, turbidity, H2S, fluorometer

Documents

Brochure
datasheet-ctd48mc-1-2-opt.pdf
665.67 KB
datasheet-ctd48mc-1-2-opt.pdf
Other documents
  • CTD Probes (5)
    754.07 KB

Ask a question

We are happy to assist.

Daniƫl Haanskorf Peter Tapken Sjors Kloosterman

We're available on all weekdays and usually reply within 1-2 working days.

Please provide detailed information.

Let us know a little about you and your contact information.

Do you represent a company or organization?

Before we proceed, would you be interested in subscribing to our newsletter?

By completing this form, you acknowledge and agree to abide by our Terms and Conditions, Privacy Policy and Cookie Policy.

Thank you for your message! Our team tipically to respond to inquiries within 1-2 business days.